Accreditation scope for

CAL 050

Bilutstyr Nor AS

St. Olavsgate 147
9407 Harstad

Telephone: 77018700
E-mail: jannerik@bilutstyrnor.no
Internet: http://www.bilutstyrnor.no
The laboratory meets the requirements in

NS-EN ISO/IEC 17025:2005




Accreditation includes:
The administrative / geographical unit:
Bilutstyr Nor AS
St. Olavsgate 147
9407 Harstad



Field testing laboratory K05 Mechanical quantities

Brake tester - brake pressure sensor

Range
Uncertainty (CMC)
Range
1 - 12 bar
Uncertainty (CMC)
± 0,07 bar ved 1, 2, 4 bar
Range
Uncertainty (CMC)
± 0,08 bar ved 6, 8, 12 bar

Remarks/ref. to internal procedure: Calibration using traceable digital manometer, stepwise applied reference pressure with manual pump. Method: Landax ID-282.

Retardation measurement

Range
Uncertainty (CMC)
Range
10,2 %G
Uncertainty (CMC)
± 2,0 %G
Range
20,8 %G
Uncertainty (CMC)
± 2,0 %G
Range
32,1 %G
Uncertainty (CMC)
± 2,0 %G
Range
44,6 %G
Uncertainty (CMC)
± 2,0 %G
Range
59,2 %G
Uncertainty (CMC)
± 2,0 %G
Range
77,3 %G
Uncertainty (CMC)
± 2,0 %G

Remarks/ref. to internal procedure: Limited to analogue instruments. Internal method 778.

Brake tester - Dead weight

Range
Uncertainty (CMC)
Range
0,01 - 30,00 kN
Uncertainty (CMC)
± 0,03 kN ved 2 kN
Range
Uncertainty (CMC)
± 0,09 kN ved 8 kN
Range
Uncertainty (CMC)
± 0,35 kN ved 30 kN

Remarks/ref. to internal procedure: Internal method 281.

Brake tester - Electronic transducer

Range
Uncertainty (CMC)
Range
0,01 - 40 kN
Uncertainty (CMC)
± 0,03 kN ved 2 kN
Range
Uncertainty (CMC)
± 0,14 kN ved 12 kN
Range
Uncertainty (CMC)
± 0,45 kN ved 40 kN

Remarks/ref. to internal procedure: Internal method 281.
Field testing laboratory K07 Fluid quantities

Diesel smoke meter - K-value

Range
Uncertainty (CMC)
Range
0,5 m-1 til 4,0 m-1
Uncertainty (CMC)
± 0,05 m-1

Remarks/ref. to internal procedure: Calibration with two or more filters. Internal method 285.

Diesel smoke meter - Opacity

Range
Uncertainty (CMC)
Range
15 % til 60 %
Uncertainty (CMC)
± 0,87 %

Remarks/ref. to internal procedure: Calibration with two or more filters. Internal method 285.

Exhaust gases analyser - Hydrocarbons

Range
Uncertainty (CMC)
Range
350 - 3500 vol-ppm
Uncertainty (CMC)
± 8 vol-ppm ved 397 vol-ppm
Range
Uncertainty (CMC)
± 51 vol-ppm ved 3200 vol-ppm

Remarks/ref. to internal procedure: 2 reference gases: Low and high range (HC = propane * PEF factor) Method: Landax ID-280. Uncertainty budget: Landax ID-304.

Exhaust gases analyser - Carbon dioxide

Range
Uncertainty (CMC)
Range
5,50 - 15,00 vol%
Uncertainty (CMC)
± 0,21 vol% ved 6,00 vol%
Range
Uncertainty (CMC)
± 0,25 vol% ved 14,00 vol%

Remarks/ref. to internal procedure: 2 reference gases: Low and high range. Method: Landax ID-280. Uncertainty budget: Landax ID-304.

Exhaust gases analyser - Carbon monoxide

Range
Uncertainty (CMC)
Range
0,40 - 5,00 vol%
Uncertainty (CMC)
± 0,02 vol% ved 0,50 vol%
Range
Uncertainty (CMC)
± 0,07 vol% ved 5,00 vol%

Remarks/ref. to internal procedure: 2 reference gases: Low and high range. Method: Landax ID-280. Uncertainty budget: Landax ID-304.
Field testing laboratory K08 Optical quantities

Headlamp beam tester

Range
Uncertainty (CMC)
Range
± 40,0 mm/m
Uncertainty (CMC)
± 1,50 mm/m

Remarks/ref. to internal procedure: Internal method 283.