You are here:
Norwegian accreditation

Accreditation scope for

CAL 044

Forsvarets laboratorietjeneste

kalibreringslaboratorium

Postboks 10
2027 Kjeller

Telephone: 63 80 87 53
E-mail: tsveen@mil.no
Internet: http://
The laboratory meets the requirements in

NS-EN ISO/IEC 17025:2017




Accreditation includes:
The administrative / geographical unit:
kalibreringslaboratorium
Fetveien 80-84
2027 Kjeller



Permanent facility K01 Electrical quantities

Electrical voltage

Range
Uncertainty (CMC)
Range
10 V
Uncertainty (CMC)
5 µV

Remarks/ref. to internal procedure: VLMS-111 (inter comparison in groups of 4)

DC-resistance

Range
Uncertainty (CMC)
Range
100 µΩ til < 1 mΩ
Uncertainty (CMC)
3,9 nΩ - 39 nΩ
Range
1 mΩ til < 10 mΩ
Uncertainty (CMC)
12 nΩ - 120 nΩ
Range
10 mΩ til < 100 mΩ
Uncertainty (CMC)
120 nΩ - 1,2 µΩ
Range
100 mΩ til 1 Ω
Uncertainty (CMC)
1,2 µΩ - 12 µΩ

Remarks/ref. to internal procedure: VLMS-133 (Calibration of resistance standards 100 µΩ - 1 Ω)

DC-resistance

Range
Uncertainty (CMC)
Range
100 mΩ-<1Ω
Uncertainty (CMC)
0,41 µΩ- 4,1 µΩ
Range
1Ω
Uncertainty (CMC)
0,81 µΩ
Range
>1Ω-10Ω
Uncertainty (CMC)
0,87 µΩ- 8,7 µΩ
Range
>10Ω-<100Ω
Uncertainty (CMC)
10 µΩ- 100 µΩ
Range
100Ω-<1kΩ
Uncertainty (CMC)
91 µΩ- 910 µΩ
Range
1kΩ-<10kΩ
Uncertainty (CMC)
0,78 mΩ-7,8 mΩ
Range
10kΩ
Uncertainty (CMC)
7,7 mΩ

Remarks/ref. to internal procedure: VLMS-134 (in oil bath, L&N 40XX or equivalent)

DC-resistance

Range
Uncertainty (CMC)
Range
100 mΩ-<1Ω
Uncertainty (CMC)
4,6 µΩ-46 µΩ
Range
1Ω
Uncertainty (CMC)
10 µΩ
Range
>1 Ω - 10 Ω
Uncertainty (CMC)
1,4 µΩ-14 µΩ
Range
>10 Ω - <100 Ω
Uncertainty (CMC)
21 µΩ-210 µΩ
Range
100 Ω - <1 kΩ
Uncertainty (CMC)
0,20 mΩ-2,0 mΩ
Range
1 kΩ -<10 kΩ
Uncertainty (CMC)
2,0 mΩ-20 mΩ
Range
10 kΩ
Uncertainty (CMC)
7,7 mΩ

Remarks/ref. to internal procedure: VLMS-134 (in air, ESI SR10XX or equivalent)

DC-resistance

Range
Uncertainty (CMC)
Range
1Ω
Uncertainty (CMC)
0,68 µΩ
Range
10 kΩ
Uncertainty (CMC)
6,4 mΩ

Remarks/ref. to internal procedure: VLMS-116 (inter comparison of 1 Ω resistance standards in groups of 4) VLMS-109 (inter comparison of 10 kΩ resistance standards in groups of 4)

DC-resistance

Range
Uncertainty (CMC)
Range
10 kΩ - 100 kΩ
Uncertainty (CMC)
30 mΩ - 300 mΩ
Range
100 kΩ - 1 MΩ
Uncertainty (CMC)
0,4 Ω - 4 Ω
Range
1 MΩ - 10 MΩ
Uncertainty (CMC)
6 Ω - 60 Ω
Range
10 MΩ - 100 MΩ
Uncertainty (CMC)
100 Ω - 1 kΩ
Range
100 MΩ - 1 GΩ
Uncertainty (CMC)
5 kΩ - 50 kΩ

Remarks/ref. to internal procedure: VLMS-135 (Calibration of resistance standards 10 kΩ - 1 GΩ)
Permanent facility K04 Dimension

Length-gauge blocks mm

Range
Uncertainty (CMC)
Range
0,5 mm - 10 mm
Uncertainty (CMC)
0,09 µm
Range
10 mm - 25 mm
Uncertainty (CMC)
0,10 µm
Range
25 mm - 50 mm
Uncertainty (CMC)
0,12 µm
Range
50 mm - 75 mm
Uncertainty (CMC)
0,14 µm
Range
75 mm - 100 mm
Uncertainty (CMC)
0,17 µm

Remarks/ref. to internal procedure: VLMM-001 (inter comparisons of gauge blocks in steel or hard metal)

Length-gauge blocks inch

Range
Uncertainty (CMC)
Range
0,02 - 0,4 in
Uncertainty (CMC)
3,5 µin
Range
0,4 - 1,0 in
Uncertainty (CMC)
3,8 µin
Range
1,0 - 2,0 in
Uncertainty (CMC)
4,6 µin
Range
2,0 - 3,0 in
Uncertainty (CMC)
5,6 µin
Range
3,0 - 4,0 in
Uncertainty (CMC)
6,8 µin

Remarks/ref. to internal procedure: VLMM-001 (inter comparisons of gauge blocks in steel or hard metal)
Field testing laboratory K05 Mechanical quantities

Pressure (gauge)

Range
Uncertainty (CMC)
Range
(-0,1 - 0,2) MPa
Uncertainty (CMC)
(0,044 - 0,050) kPa
Range
(0,2 - 2) MPa
Uncertainty (CMC)
(0,37 - 0,44) kPa
Range
(2 - 20) MPa
Uncertainty (CMC)
(3,7 - 4,3) kPa

Remarks/ref. to internal procedure: VLMM-025 (Calibration of pressure gauge) - Field laboratory
Permanent facility K05 Mechanical quantities

Pressure (gauge)

Range
Uncertainty (CMC)
Range
(-0,1 - 0,2) MPa
Uncertainty (CMC)
(0,044 - 0,050) kPa
Range
(0,2 - 2) MPa
Uncertainty (CMC)
(0,37 - 0,44) kPa
Range
(2 - 20) MPa
Uncertainty (CMC)
(3,7 - 4,3) kPa

Remarks/ref. to internal procedure: VLMM-025 (Calibration of pressure gauge) - Permanent laboratory