Diesel smoke meter - K-value
Range
0,7 m-1 - 7 m-1
Uncertainty (CMC)
± 0,05 m-1 ved 0,7 m-1
Range
Uncertainty (CMC)
± 0,06 m-1 ved 7 m-1
Remarks/ref. to internal procedure: Calibration with two or more filters. Procedure dok id 308.
Diesel smoke meter - Opacity
Range
12 % - 75,5 %
Uncertainty (CMC)
± 0,7 % ved 12 %
Range
Uncertainty (CMC)
± 0,7 % ved 75,5 %
Remarks/ref. to internal procedure: Calibration with two or more filters. Procedure dok id 308.
Exhaust gases analyser - Carbon monoxide
Range
0,48 vol% - 4,20 vol%
Uncertainty (CMC)
± 0,03 vol% ved 0,48 vol%
Range
Uncertainty (CMC)
± 0,05 vol% ved 4,20 vol%
Remarks/ref. to internal procedure: Minimum 2 reference gases. Procedure dok id 297
Exhaust gases analyser - Carbon dioxide
Range
3,4 vol% - 14,25 vol%
Uncertainty (CMC)
± 0,1 vol% ved 3,4 vol%
Range
Uncertainty (CMC)
± 0,2 vol% ved 14,25 vol%
Remarks/ref. to internal procedure: Minimum 2 reference gases. Procedure dok id 297
Exhaust gases analyser - Hydrocarbons.
Range
140 vol ppm (heksan) - 2000 vol ppm (heksan)
Uncertainty (CMC)
± 5 vol ppm ved 140 vol ppm (*)
* referansegass leveres med 1% usikkerhet
Range
Uncertainty (CMC)
± 7 vol ppm ved 500 vol ppm (**)
** referansegass leveres med 2% usikkerhet
Range
Uncertainty (CMC)
± 13 vol ppm ved 1100 vol ppm (*)
Range
Uncertainty (CMC)
± 23 vol ppm ved 2000 vol ppm (*)
Remarks/ref. to internal procedure: Minimum 2 reference gases. Propan is used. The objects PEF value are the conversion factor between hexane/propane. Procedure dok id 297