Accreditation scope for

CAL 009

Trescal Norway AS

Postboks 124
4098 TANANGER

Telephone: Tlf. 51 71 95 00 Faks. 51 71 95 01
E-mail: contact.norway@trescal.com
Internet: http://www.trescal.no
The laboratory meets the requirements in

NS-EN ISO/IEC 17025:2017




Accreditation includes:
The administrative / geographical unit:
IKM Laboratorium AS avd Tananger
NorSea Base
Bygg 53
4056 Tananger



Permanent facility K01 Electrical quantities

DC power, calibration of meter

Range
Uncertainty (CMC)
Range
1 to 100 mW
Uncertainty (CMC)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (CMC)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (CMC)
0,064 - 0,16 %

Remarks/ref. to internal procedure: BL08.04.01

Capacitance, calibration of meter

Range
Uncertainty (CMC)
Range
0,19 to 1,1 nF
Uncertainty (CMC)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (CMC)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (CMC)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (CMC)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (CMC)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (CMC)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (CMC)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (CMC)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (CMC)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (CMC)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (CMC)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (CMC)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (CMC)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (CMC)
0,65 - 0,92 %, 0-0.2 Hz

Remarks/ref. to internal procedure: BL08.04.01

Resistance, measurement

Range
Uncertainty (CMC)
Range
1 to 100 mohm
Uncertainty (CMC)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (CMC)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (CMC)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (CMC)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (CMC)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (CMC)
0,024 - 2 kohm

Remarks/ref. to internal procedure: BL08.06.01

Resistance, calibration of meter

Range
Uncertainty (CMC)
Range
0 ohm
Uncertainty (CMC)
39 µohm
Range
0 - 1 ohm
Uncertainty (CMC)
85 µohm
Range
1 - 1,9 ohm
Uncertainty (CMC)
160 µohm
Range
1,9 - 10 ohm
Uncertainty (CMC)
250 µohm
Range
10 - 25ohm
Uncertainty (CMC)
74 µohm
Range
25 - 19ohm
Uncertainty (CMC)
450 µohm
Range
19 - 100ohm
Uncertainty (CMC)
180 µohm
Range
100 - 190ohm
Uncertainty (CMC)
2,9 mohm
Range
190 - 1kohm
Uncertainty (CMC)
16 mohm
Range
1 - 1,9kohm
Uncertainty (CMC)
30 mohm
Range
1,9 - 10kohm
Uncertainty (CMC)
160 mohm
Range
10 - 19kohm
Uncertainty (CMC)
300 mohm
Range
19 - 100kohm
Uncertainty (CMC)
1,6 ohm
Range
100 - 1Mohm
Uncertainty (CMC)
3,0 ohm
Range
1 - 1,9Mohm
Uncertainty (CMC)
35 ohm
Range
1,9 - 10Mohm
Uncertainty (CMC)
350 ohm
Range
10 - 19Mohm
Uncertainty (CMC)
810 ohm
Range
19 - 100Mohm
Uncertainty (CMC)
10 kohm
Range
0,1 - 1,1Gohm
Uncertainty (CMC)
0,4 - 12 %

Remarks/ref. to internal procedure: BL08.04.01

AC Voltage, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 2,2 mV
Uncertainty (CMC)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (CMC)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (CMC)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (CMC)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (CMC)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (CMC)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (CMC)
19 - 91 mV, 40-01k Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Voltage, calibration of source

Range
Uncertainty (CMC)
Range
0 to 120 mV
Uncertainty (CMC)
1 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (CMC)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (CMC)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (CMC)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (CMC)
0,82 - 6,1 mV

Remarks/ref. to internal procedure: BL08.05.01

DC Resistance, calibration of source

Range
Uncertainty (CMC)
Range
0 to 12 ohm
Uncertainty (CMC)
0,05 - 0,23 mohm
BL08.06.03
Range
12 to 120 ohm
Uncertainty (CMC)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (CMC)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (CMC)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (CMC)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (CMC)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (CMC)
160 - 700 ohm

Remarks/ref. to internal procedure: BL08.06.03

DC Current

Range
Uncertainty (CMC)
Range
0 to 120 µA
Uncertainty (CMC)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (CMC)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (CMC)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (CMC)
0,9 - 4,7 µA
Range
0,12 to 1,0 A
Uncertainty (CMC)
0,023 - 0,12mA
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage

Range
Uncertainty (CMC)
Range
0 to 120 mV
Uncertainty (CMC)
1,0 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (CMC)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (CMC)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (CMC)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (CMC)
0,82 - 6,1 mV
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: BL08.05.01

DC Resistance

Range
Uncertainty (CMC)
Range
0 to 12 ohm
Uncertainty (CMC)
0,05 - 0,23 mohm
Range
12 to 120 ohm
Uncertainty (CMC)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (CMC)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (CMC)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (CMC)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (CMC)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (CMC)
160 - 700 ohm
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: BL08.06.03

DC Voltage, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 mV
Uncertainty (CMC)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (CMC)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (CMC)
17 - 71 µV
Range
11 to 22 V
Uncertainty (CMC)
74 - 140 µV
Range
22 to 220 V
Uncertainty (CMC)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (CMC)
2,3 - 9,8 mV

Remarks/ref. to internal procedure: BL08.04.01

AC Current, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 µA
Uncertainty (CMC)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (CMC)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (CMC)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (CMC)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (CMC)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (CMC)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (CMC)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (CMC)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (CMC)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (CMC)
2,6 - 4,8 A, 45-100 Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Current, calibration of source

Range
Uncertainty (CMC)
Range
0 to 120 µA
Uncertainty (CMC)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (CMC)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (CMC)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (CMC)
0,90 - 4,7 µA
Range
0,12 to 1 A
Uncertainty (CMC)
0,023 - 0,12 mA

Remarks/ref. to internal procedure: BL08.05.01

DC Current, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 µA
Uncertainty (CMC)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (CMC)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (CMC)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (CMC)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (CMC)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (CMC)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (CMC)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (CMC)
0,092 - 4,8 A

Remarks/ref. to internal procedure: BL08.04.01

Calibration of resistance bridge

Range
Uncertainty (CMC)
Range
0,25 to 4 ohm/ohm
Uncertainty (CMC)
2 - 17 µohm/ohm

Remarks/ref. to internal procedure: BL08.06.02

Termocouple (TC) Measure, cal of source

Range
Uncertainty (CMC)
Range
-250 °C to 2316 °C
Uncertainty (CMC)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.05.01

Resistance thermometer Source, cal of meter

Range
Uncertainty (CMC)
Range
-200 °C to 630 °C
Uncertainty (CMC)
From U=0,02 °C to 0,18 °C

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Source, cal of meter

Range
Uncertainty (CMC)
Range
-250 °C to 2316 °C
Uncertainty (CMC)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.04.01
Permanent facility K03 Time and frequency

Periode, periode time

Range
Uncertainty (CMC)
Range
0,1µs to 100ms
Uncertainty (CMC)
1,0E-9 * Ptime+trigger error

Remarks/ref. to internal procedure: BL21.04.05

Number, number of pulses

Range
Uncertainty (CMC)
Range
1 to 10G
Uncertainty (CMC)
0,60

Remarks/ref. to internal procedure: BL21.04.06

Frequency, as measured quantity

Range
Uncertainty (CMC)
Range
0,5Hz to 5GHz
Uncertainty (CMC)
1,0E-9*f

Remarks/ref. to internal procedure: BL21.04.03

Frequency, as source

Range
Uncertainty (CMC)
Range
1,5 and 10MHz
Uncertainty (CMC)
1,0E-9 * f

Remarks/ref. to internal procedure: BL21.04.03

Time interval

Range
Uncertainty (CMC)
Range
1 to 200 hour
Uncertainty (CMC)
0,80s

Remarks/ref. to internal procedure: BL21.04.04

Frequency, calibration of meter

Range
Uncertainty (CMC)
Range
0,010Hz to 2,0MHz
Uncertainty (CMC)
5,0 µHz + 1,7 ppm

Remarks/ref. to internal procedure: BL08.04.01

Frequency, calibration of source

Range
Uncertainty (CMC)
Range
3Hz to 1MHz
Uncertainty (CMC)
50 µHz + 50 ppm

Remarks/ref. to internal procedure: BL08.05.01
Permanent facility K04 Dimension

Thread gauge, plug, simple pitch diameter

Range
Uncertainty (CMC)
Range
2,2 to 100 mm
Uncertainty (CMC)
3,5 µm
Range
100 to 150 mm
Uncertainty (CMC)
4,0 µm

Remarks/ref. to internal procedure: BL14.05.03

Thread gauge, ring, simple pitch diameter

Range
Uncertainty (CMC)
Range
2,2 to 100 mm
Uncertainty (CMC)
5,0 µm
Range
100 to 150 mm
Uncertainty (CMC)
5,4 µm

Remarks/ref. to internal procedure: BL14.05.04

Caliper

Range
Uncertainty (CMC)
Range
0 to 1000 mm
Uncertainty (CMC)
14 µm + 1,6E-5*L

Remarks/ref. to internal procedure: BL14.04.01 (L = measured length). Depth measurement is not covered by scope of accreditation.

Dial gauge

Range
Uncertainty (CMC)
Range
0 to 1 mm
Uncertainty (CMC)
0,5 µm
At resolution: 0.001 mm
Range
0 to 100 mm
Uncertainty (CMC)
1,2 µm + 1,1E-4*L

Remarks/ref. to internal procedure: BL14.04.04

Gauge block, steel

Range
Uncertainty (CMC)
Range
0,5 to 100 mm
Uncertainty (CMC)
0,13 µm +1,2E-6*L

Remarks/ref. to internal procedure: BL14.07.01

Length, external

Range
Uncertainty (CMC)
Range
0 to 600 mm
Uncertainty (CMC)
0,55 µm + 3,0E-6*L
Range
600 to 1000 mm
Uncertainty (CMC)
2,1 µm + 3,5E-6*L

Remarks/ref. to internal procedure: BL14.05.01

Micrometer, external

Range
Uncertainty (CMC)
Range
0 to 500 mm
Uncertainty (CMC)
1,1 µm + 2,1E-05*L

Remarks/ref. to internal procedure: BL14.04.02
Permanent facility K05 Mechanical quantities

Weight/Mass

Range
Uncertainty (CMC)
Range
0,1 g to 120 g
Uncertainty (CMC)
± 0,26 mg + 1,0 mg/kg
Range
120 g to 2300 g
Uncertainty (CMC)
± 0,60 mg + 2,0 mg/kg
Range
2,3 kg to 20 kg
Uncertainty (CMC)
± 10 mg + 3,0 mg/kg

Remarks/ref. to internal procedure: BL06.07.01

Pressure (abs)

Range
Uncertainty (CMC)
Range
20 to 100 hPa
Uncertainty (CMC)
2,4 Pa + 200 ppm
Range
0,01 to 7,6 Mpa
Uncertainty (CMC)
0,20 Pa + 30 ppm
Range
7,6 to 100 Mpa
Uncertainty (CMC)
75 Pa + 34 ppm
Range
100 to 200 Mpa
Uncertainty (CMC)
75 Pa + 39 ppm
Range
200 to 350 Mpa
Uncertainty (CMC)
1500 Pa + 150 ppm

Remarks/ref. to internal procedure: BL06.05.01/ 06.08.01

Pressure (diff)

Range
Uncertainty (CMC)
Range
0 to 0,5 Mpa
Uncertainty (CMC)
8,0 - 58 Pa

Remarks/ref. to internal procedure: BL06.05.02/ 06.08.02

Dividing ratio -oil and gas

Range
Uncertainty (CMC)
Range
0,01 to 10 Mpa
Uncertainty (CMC)
50 ppm

Remarks/ref. to internal procedure: BL06.04.04

Kn value -oil and gas

Range
Uncertainty (CMC)
Range
0,01 to 7,6 Mpa
Uncertainty (CMC)
35 ppm

Remarks/ref. to internal procedure: BL06.04.06

Piston/sylinder area -oil and gas

Range
Uncertainty (CMC)
Range
0,01 to 7,6 Mpa
Uncertainty (CMC)
30 ppm
Range
7,6 to 100 Mpa
Uncertainty (CMC)
30 ppm
Range
100 to 200 Mpa
Uncertainty (CMC)
34 ppm

Remarks/ref. to internal procedure: BL06.04.01/06.04.05

Pressure (gauge)

Range
Uncertainty (CMC)
Range
-0,095 to 0 Mpa
Uncertainty (CMC)
20 Pa
Range
0 to 100 hPa
Uncertainty (CMC)
0,3 Pa + 75 ppm
Range
0,01 to 7,6 Mpa
Uncertainty (CMC)
0,20 Pa + 30 ppm
Range
7,6 to 100 Mpa
Uncertainty (CMC)
75 Pa + 34 ppm
Range
100 to 200 Mpa
Uncertainty (CMC)
75 Pa + 39 ppm
Range
200 to 350 Mpa
Uncertainty (CMC)
1500 Pa + 150 ppm

Remarks/ref. to internal procedure: BL06.05.01/ 06.08.01
Permanent facility K10 Temperature, humidity and thermophysical properties

Resistance termometer

Range
Uncertainty (CMC)
Range
-80 to 300 °C
Uncertainty (CMC)
18 mK +|3 mK/100°C|

Remarks/ref. to internal procedure: BL07.04.02

Indicator

Range
Uncertainty (CMC)
Range
-80 to 300 °C
Uncertainty (CMC)
18 mK +|3 mK/100°C|
Range
300 to 660 °C
Uncertainty (CMC)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.06.01

Transmitter

Range
Uncertainty (CMC)
Range
-80 to 300 °C
Uncertainty (CMC)
18 mK +|3 mK/100°C|
Range
300 to 660 °C
Uncertainty (CMC)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.05.01

Dry block calibrator

Range
Uncertainty (CMC)
Range
-80 to 660°C
Uncertainty (CMC)
50-300 mK

Remarks/ref. to internal procedure: BL07.08.03
The administrative / geographical unit:
IKM Laboratorium AS avd Tananger field calibration
NorSea Base
Bygg 53
4056 Tananger



Field testing laboratory K01 Electrical quantities

DC power, calibration of meter

Range
Uncertainty (CMC)
Range
1 to 100 mW
Uncertainty (CMC)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (CMC)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (CMC)
0,064 - 0,16 %

Remarks/ref. to internal procedure: BL08.04.01

Capacitance, calibration of meter

Range
Uncertainty (CMC)
Range
0,19 to 1,1 nF
Uncertainty (CMC)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (CMC)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (CMC)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (CMC)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (CMC)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (CMC)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (CMC)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (CMC)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (CMC)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (CMC)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (CMC)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (CMC)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (CMC)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (CMC)
0,65 - 0,92 %, 0-0.2 Hz

Remarks/ref. to internal procedure: BL08.04.01

Resistance, measurment

Range
Uncertainty (CMC)
Range
1 to 100 mohm
Uncertainty (CMC)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (CMC)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (CMC)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (CMC)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (CMC)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (CMC)
0,024 - 2 kohm

Remarks/ref. to internal procedure: BL08.06.01

Resistance, calibration of meter

Range
Uncertainty (CMC)
Range
0 ohm
Uncertainty (CMC)
39 µohm
Range
0 to 1 ohm
Uncertainty (CMC)
85 µohm
Range
1 to 1,9 ohm
Uncertainty (CMC)
160 µohm
Range
1,9 to 10 ohm
Uncertainty (CMC)
250 µohm
Range
10 to 25 ohm
Uncertainty (CMC)
74 µohm
Range
25 to 19 ohm
Uncertainty (CMC)
450 µohm
Range
19 to 100 ohm
Uncertainty (CMC)
180 µohm
Range
100 to 190 ohm
Uncertainty (CMC)
2,9 mohm
Range
190 to 1 kohm
Uncertainty (CMC)
16 mohm
Range
1 to 1,9 kohm
Uncertainty (CMC)
30 mohm
Range
1,9 to 10 kohm
Uncertainty (CMC)
160 mohm
Range
10 to 19 kohm
Uncertainty (CMC)
300 mohm
Range
19 to 100 kohm
Uncertainty (CMC)
1,6 ohm
Range
100 to 1 Mohm
Uncertainty (CMC)
3,0 ohm
Range
1 to 1,9 Mohm
Uncertainty (CMC)
35 ohm
Range
1,9 to 10 Mohm
Uncertainty (CMC)
350 ohm
Range
10 to 19 Mohm
Uncertainty (CMC)
810 ohm
Range
19 to 100 Mohm
Uncertainty (CMC)
10 kohm
Range
0,1 to 1,1 Gohm
Uncertainty (CMC)
0,4 - 12 %

Remarks/ref. to internal procedure: BL08.04.01

AC Voltage, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 2,2 mV
Uncertainty (CMC)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (CMC)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (CMC)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (CMC)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (CMC)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (CMC)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (CMC)
19 - 91 mV, 40-01k Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Voltage, calibration of source

Range
Uncertainty (CMC)
Range
0 to 100 mV
Uncertainty (CMC)
2,7 - 5,6 µV
Range
0,1 to 1 V
Uncertainty (CMC)
7,4 - 25 µV
Range
1 to 10 V
Uncertainty (CMC)
57 - 220 µV
Range
10 to 100 V
Uncertainty (CMC)
0,76 - 3,4 mV
Range
100 to 1000 V
Uncertainty (CMC)
11 - 40 mV

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 mV
Uncertainty (CMC)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (CMC)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (CMC)
17 - 71 µV
Range
11 to 22 V
Uncertainty (CMC)
74 - 140 µV
Range
22 to 220 V
Uncertainty (CMC)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (CMC)
2,3 - 9,8 mV

Remarks/ref. to internal procedure: BL08.04.01

AC Current, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 µA
Uncertainty (CMC)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (CMC)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (CMC)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (CMC)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (CMC)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (CMC)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (CMC)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (CMC)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (CMC)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (CMC)
2,6 - 4,8 A, 45-100 Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Current, calibration of source

Range
Uncertainty (CMC)
Range
0 to 10 mA
Uncertainty (CMC)
50 - 250 nA
Range
10 to 30 mA
Uncertainty (CMC)
0,73 - 1,1 µA
Range
30 to 100 mA
Uncertainty (CMC)
15 - 41 µA

Remarks/ref. to internal procedure: BL08.05.01

DC Current, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 µA
Uncertainty (CMC)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (CMC)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (CMC)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (CMC)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (CMC)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (CMC)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (CMC)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (CMC)
0,092 - 4,8 A

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Measure, cal of source

Range
Uncertainty (CMC)
Range
-250 °C to 2316 °C
Uncertainty (CMC)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.05.01

Resistance thermometer Source, cal of meter

Range
Uncertainty (CMC)
Range
-200 °C to 630 °C
Uncertainty (CMC)
From U=0,02 °C to 0,18 °C

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Source, cal of meter

Range
Uncertainty (CMC)
Range
-250 °C to 2316 °C
Uncertainty (CMC)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.04.01
Field testing laboratory K05 Mechanical quantities

Balances

Range
Uncertainty (CMC)
Range
0,1 to 120 g
Uncertainty (CMC)
0,16-0,52 mg
Range
120 to 5000 g
Uncertainty (CMC)
2,0-17 mg
Range
5 to 16 kg
Uncertainty (CMC)
31-64 mg
Range
16 to 60 kg
Uncertainty (CMC)
0,48-1,4 g

Remarks/ref. to internal procedure: BL06.07.02

Pressure (gauge)

Range
Uncertainty (CMC)
Range
-0,095 to 0 Mpa
Uncertainty (CMC)
73 - 40 Pa
Range
0 to 0,35MPa
Uncertainty (CMC)
1,7 - 43 Pa
Range
0,35 to 6MPa
Uncertainty (CMC)
0,72 - 7,5 hPa
Range
6 to 350MPa
Uncertainty (CMC)
0,75 - 53 kPa

Remarks/ref. to internal procedure: BL06.05.01/ 06.08.01
Field testing laboratory K10 Temperature, humidity and thermophysical properties

Indicator

Range
Uncertainty (CMC)
Range
-25 to 100 °C
Uncertainty (CMC)
80 mK
Range
100 to 200 °C
Uncertainty (CMC)
100 mK
Range
200 to 660 °C
Uncertainty (CMC)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.06.01

Transmitter

Range
Uncertainty (CMC)
Range
-25 to 100 °C
Uncertainty (CMC)
80 mK
Range
100 to 200 °C
Uncertainty (CMC)
100 mK
Range
200 to 660 °C
Uncertainty (CMC)
50 mK + |50 mK/100°C|

Remarks/ref. to internal procedure: BL07.05.01

Calibration of ovens and furnaces (Sensor Accuracy Test)

Range
Uncertainty (CMC)
Range
50 til 650 °C
Uncertainty (CMC)
±0,60 til ±0,90 °C

Remarks/ref. to internal procedure: BL07.08.04

Calibration of ovens and furnaces (Temperature Uniformity Survey)

Range
Uncertainty (CMC)
Range
50 til 650 °C
Uncertainty (CMC)
±0,60 til ±0,90 °C

Remarks/ref. to internal procedure: BL07.08.01
The administrative / geographical unit:
IKM Laboratorium avd Oslo
Ensjøveien 18
0661 Oslo



Permanent facility K01 Electrical quantities

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,040 %

Remarks/ref. to internal procedure: BL08.07.03 Based on "Statnetts krav til måling av sentralnettsutveksling v.1.1(2016)" and includes evaluation of metering point. cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V

DC power, calibration of meter

Range
Uncertainty (CMC)
Range
1 to 100 mW
Uncertainty (CMC)
0,029 - 0,026 %
Range
0,1 to 100 W
Uncertainty (CMC)
0,026 - 0,064 %
Range
0,1 to 21 kW
Uncertainty (CMC)
0,064 - 0,16 %

Remarks/ref. to internal procedure: BL08.04.01

Capacitance, calibration of meter

Range
Uncertainty (CMC)
Range
0,19 to 1,1 nF
Uncertainty (CMC)
4,5 - 1,1 %, 10-10k Hz
Range
1,1 to 3,3 nF
Uncertainty (CMC)
1,1 - 0,62 %, 10-3k Hz
Range
3,3 to 330 nF
Uncertainty (CMC)
0,62 - 0,26 %, 10-1k Hz
Range
0,33 to 1,1 µF
Uncertainty (CMC)
0,26 %, 10-600 Hz
Range
1,1 to 3,3 µF
Uncertainty (CMC)
0,26 %, 10-300 Hz
Range
3,3 to 11 µF
Uncertainty (CMC)
0,26 %, 10-150 Hz
Range
11 to 33 µF
Uncertainty (CMC)
0,26 - 0,38%, 10-120 Hz
Range
33 to 110 µF
Uncertainty (CMC)
0,38 - 0,42%, 10-80 Hz
Range
110 to 330 µF
Uncertainty (CMC)
0,42 %, 0-50 Hz
Range
0,33 to 1,1 mF
Uncertainty (CMC)
0,42 %, 0-20 Hz
Range
1,1 to 3,3 mF
Uncertainty (CMC)
0,42 %, 0-6 Hz
Range
3,3 to 11 mF
Uncertainty (CMC)
0,42 %, 0-2 Hz
Range
11 to 33 mF
Uncertainty (CMC)
0,42 - 0,65 %, 0-0.6 Hz
Range
33 to 110 mF
Uncertainty (CMC)
0,65 - 0,92 %, 0-0.2 Hz

Remarks/ref. to internal procedure: BL08.04.01

Resistance, measurement

Range
Uncertainty (CMC)
Range
1 to 100 mohm
Uncertainty (CMC)
0,47 - 11 µohm
Range
0,1 to 17,4 ohm
Uncertainty (CMC)
0,011 - 1,4 mohm
Range
17,4 to 400 ohm
Uncertainty (CMC)
0,20 - 3,1 mohm
Range
0,4 to 1 kohm
Uncertainty (CMC)
29 - 72 mohm
Range
1 to 100 kohm
Uncertainty (CMC)
0,072 - 24 ohm
Range
0,1 to 1 Mohm
Uncertainty (CMC)
0,024 - 2 kohm

Remarks/ref. to internal procedure: BL08.06.01

Resistance, calibration of meter

Range
Uncertainty (CMC)
Range
0 ohm
Uncertainty (CMC)
39 µohm
Range
0 to 1 ohm
Uncertainty (CMC)
85 µohm
Range
1 to 1,9 ohm
Uncertainty (CMC)
160 µohm
Range
1,9 to 10 ohm
Uncertainty (CMC)
250 µohm
Range
10 to 25ohm
Uncertainty (CMC)
74 µohm
Range
25 to 19ohm
Uncertainty (CMC)
450 µohm
Range
19 to 100ohm
Uncertainty (CMC)
180 µohm
Range
100 to 190ohm
Uncertainty (CMC)
2,9 mohm
Range
190 to 1kohm
Uncertainty (CMC)
16 mohm
Range
1 to 1,9kohm
Uncertainty (CMC)
30 mohm
Range
1,9 to 10kohm
Uncertainty (CMC)
160 mohm
Range
10 to 19kohm
Uncertainty (CMC)
300 mohm
Range
19 to 100kohm
Uncertainty (CMC)
1,6 ohm
Range
100 to 1Mohm
Uncertainty (CMC)
3,0 ohm
Range
1 to 1,9Mohm
Uncertainty (CMC)
35 ohm
Range
1,9 to 10Mohm
Uncertainty (CMC)
350 ohm
Range
10 to 19Mohm
Uncertainty (CMC)
810 ohm
Range
19 to 100Mohm
Uncertainty (CMC)
10 kohm
Range
0,1 to 1,1Gohm
Uncertainty (CMC)
0,4 - 12 %

Remarks/ref. to internal procedure: BL08.04.01

AC Voltage, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 2,2 Mv
Uncertainty (CMC)
7,8 - 8,2 µV, 40-20k Hz
Range
2,2 to 22 mV
Uncertainty (CMC)
9,7 - 13 µV, 40-20k Hz
Range
22 to 220 mV
Uncertainty (CMC)
9,6 - 27 µV, 40-20k Hz
Range
0,22 to 2,2 V
Uncertainty (CMC)
20 - 150 µV, 40-20k Hz
Range
2,2 to 22 V
Uncertainty (CMC)
0,2 - 1,5 mV, 40-20k Hz
Range
22 to 220 V
Uncertainty (CMC)
2,3 - 16 mV, 40-20k Hz
Range
220 to 1100 V
Uncertainty (CMC)
19 - 91 mV, 40-01k Hz

Remarks/ref. to internal procedure: BL08.04.01

DC Current

Range
Uncertainty (CMC)
Range
0 to 120 µA
Uncertainty (CMC)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (CMC)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (CMC)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (CMC)
0,9 - 4,7 µA
Range
0,12 to 1,0 A
Uncertainty (CMC)
0,023 - 0,12mA
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage

Range
Uncertainty (CMC)
Range
0 to 120 mV
Uncertainty (CMC)
1,0 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (CMC)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (CMC)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (CMC)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (CMC)
0,82 - 6,1 mV
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: BL08.05.01

DC Resistance

Range
Uncertainty (CMC)
Range
0 to 12 ohm
Uncertainty (CMC)
0,05 - 0,23 mohm
Range
12 to 120 ohm
Uncertainty (CMC)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (CMC)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (CMC)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (CMC)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (CMC)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (CMC)
160 - 700 ohm

Remarks/ref. to internal procedure: BL08.06.03

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,020 %

Remarks/ref. to internal procedure: BL08.07.04 cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V

DC Resistance, calibration of source

Range
Uncertainty (CMC)
Range
0 to 12 ohm
Uncertainty (CMC)
0,05 - 0,23 mohm
BL08.06.03
Range
12 to 120 ohm
Uncertainty (CMC)
0,64 - 1,9 mohm
Range
0,12 to 1,2 kohm
Uncertainty (CMC)
1,7 - 13 mohm
Range
1,2 to 12 kohm
Uncertainty (CMC)
0,020 - 0,13 ohm
Range
12 to 120 kohm
Uncertainty (CMC)
0,20 - 1,3 ohm
Range
0,12 to 1,2 Mohm
Uncertainty (CMC)
4,0 - 20 ohm
Range
1,2 to 12 Mohm
Uncertainty (CMC)
160 - 700 ohm
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: BL08.06.03

DC Voltage, calibration of source

Range
Uncertainty (CMC)
Range
0 to 120 mV
Uncertainty (CMC)
1 - 1,6 µV
Range
0,12 to 1,2 V
Uncertainty (CMC)
1,5 - 5,8 µV
Range
1,2 to 12 V
Uncertainty (CMC)
7,0 - 50 µV
Range
12 to 120 V
Uncertainty (CMC)
0,10 - 0,75 mV
Range
120 to 1000 V
Uncertainty (CMC)
0,82 - 6,1 mV

Remarks/ref. to internal procedure: BL08.05.01

DC Voltage, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 mV
Uncertainty (CMC)
0,68 - 2,9 µV
Range
0,22 to 2,2 V
Uncertainty (CMC)
3,0 - 15 µV
Range
2,2 to 11 V
Uncertainty (CMC)
17 - 71 µV
Range
11 to 22 V
Uncertainty (CMC)
74 - 140 µV
Range
22 to 220 V
Uncertainty (CMC)
0,23 - 1,6 mV
Range
220 to 1100 V
Uncertainty (CMC)
2,3 - 9,8 mV

Remarks/ref. to internal procedure: BL08.04.01

DC Current, calibration of source

Range
Uncertainty (CMC)
Range
0 to 120 µA
Uncertainty (CMC)
0,1 - 2,5 nA
Range
0,12 to 1,2 mA
Uncertainty (CMC)
7,4 - 29 nA
Range
1,2 to 12 mA
Uncertainty (CMC)
0,074 - 0,29 µA
Range
12 to 120 mA
Uncertainty (CMC)
0,90 - 4,7 µA
Range
0,12 to 1 A
Uncertainty (CMC)
0,023 - 0,12 mA

Remarks/ref. to internal procedure: BL08.05.01

DC Current, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 µA
Uncertainty (CMC)
7,8 - 18 nA
Range
0,22 to 2,2 mA
Uncertainty (CMC)
18 - 110 nA
Range
2,2 to 22 mA
Uncertainty (CMC)
0,18 - 1,1 µA
Range
22 to 220 mA
Uncertainty (CMC)
2,0 - 13 µA
Range
0,22 to 2,2 A
Uncertainty (CMC)
39 - 185 µA
Range
2,2 to 11 A
Uncertainty (CMC)
0,99 - 3,4 mA
Range
11 to 20,5 A
Uncertainty (CMC)
9,1 - 16 mA
Range
20,5 to 1025 A
Uncertainty (CMC)
0,092 - 4,8 A

Remarks/ref. to internal procedure: BL08.04.01

Voltage ratio

Range
Uncertainty (CMC)
Range
HF Powermeter DC verification
Uncertainty (CMC)
Range
0 to 100 % scale
Uncertainty (CMC)
0,06 - 0,6 % point
Zero carryover (0% scale) and instrumentation (100 % scale)

Remarks/ref. to internal procedure: MP 220

AC Current, calibration of meter

Range
Uncertainty (CMC)
Range
0 to 220 µA
Uncertainty (CMC)
31 - 86 nA, 40-1k Hz
Range
0,22 to 2,2 mA
Uncertainty (CMC)
0,12 - 0,61 µA, 40-1k Hz
Range
2,2 to 22 mA
Uncertainty (CMC)
0,87 - 4,5 µA, 40-1k Hz
Range
22 to 220 mA
Uncertainty (CMC)
12 - 68 µA, 40-1k Hz
Range
0,22 to 2,2 A
Uncertainty (CMC)
0,16 - 1,3 µA, 20-1k Hz
Range
2,2 to 11 A
Uncertainty (CMC)
1,3 - 6,1 µA, 40-1k Hz
Range
11 to 20,5 A
Uncertainty (CMC)
14 - 23 mA, 45-100 Hz
Range
20,5 to 110 A
Uncertainty (CMC)
93 - 500 mA, 20-1k Hz
Range
110 to 550 A
Uncertainty (CMC)
0,50 - 2,5 A, 40-1k Hz
Range
550 to 1025 A
Uncertainty (CMC)
2,6 - 4,8 A, 45-100 Hz

Remarks/ref. to internal procedure: BL08.04.01

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,10 %

Remarks/ref. to internal procedure: BL08.07.01 cosφ = -1 til 1 f = 45 Hz til 70Hz I = 0,01 A til 120 A U = 5 V til 520 V

Power, using power sensor

Range
Uncertainty (CMC)
Range
0,9 to 1,1 mW
Uncertainty (CMC)
3,1 µW
49 - 51 MHz in 50 ohm system and N connector.
Range
+20 to -70 dBm
Uncertainty (CMC)
0,01 - 0,69 dB
100 kHz to 18 GHz in 50 ohm system and N connector
Range
+20 to -30 dBm
Uncertainty (CMC)
0,01 - 1,48 dB
10 MHz to 26.5 GHz in 50 ohm system and 3.5 mm connector

Remarks/ref. to internal procedure: MP 210 Measured using Power Sensor

Attenuation

Range
Uncertainty (CMC)
Range
0 to 50 dB
Uncertainty (CMC)
0,02 - 0,46 dB
100 kHz to 18 GHz in 50 ohm system with N connector

Remarks/ref. to internal procedure: MP 212 RF substitution

Cal Factor for power sensor with N-connector

Range
Uncertainty (CMC)
Range
80 to 110 %
Uncertainty (CMC)
0,3 - 0,8 % points
100 kHz to 18 GHz in 50 ohm system and N connector
Range
80 to 110 %
Uncertainty (CMC)
Range
Uncertainty (CMC)
0,4 - 1,8 % points
10 MHz to 26,5 GHz in 50 ohm system and 3,5 mm connector

Remarks/ref. to internal procedure: MP 213

Power, generated level using power splitter

Range
Uncertainty (CMC)
Range
1 mW (0 dBm)
Uncertainty (CMC)
3,1 µW
50 MHz in 50 ohm system with N connector
Range
+20 to - 70 dBm
Uncertainty (CMC)
0,04 - 0,68 dB
100 kHz to 18 GHz in 50 ohm system and N connector. The uncertainty will be greater if adaptors must be used.
Range
+10 to - 30 dBm
Uncertainty (CMC)
0,02 - 1,48 dB
10 MHz to 26.5 GHz in 50 ohm system and 3.5 mm connector.

Remarks/ref. to internal procedure: MP 211

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,10%

Remarks/ref. to internal procedure: BL08.07.01 cosφ = -1 til 1 f = 45 Hz til 70Hz I = 0,01 A til 120 A U = 5 V til 520 V

Termocouple (TC) Source, cal of meter

Range
Uncertainty (CMC)
Range
-250 °C to 2316 °C
Uncertainty (CMC)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.04.01

Termocouple (TC) Measure, cal of source

Range
Uncertainty (CMC)
Range
-250 °C to 2316 °C
Uncertainty (CMC)
From U=0,11 °C to 0,65 °C

Remarks/ref. to internal procedure: BL08.05.01

Resistance thermometer Source, cal of meter

Range
Uncertainty (CMC)
Range
-200 °C to 630 °C
Uncertainty (CMC)
From U=0,02 °C to 0,18 °C

Remarks/ref. to internal procedure: BL08.04.01
Permanent facility K03 Time and frequency

Periode, periode time

Range
Uncertainty (CMC)
Range
0,1µs to 100ms
Uncertainty (CMC)
1,0E-9 * Ptime+trigger error

Remarks/ref. to internal procedure: BL21.04.05

Frequency, as source

Range
Uncertainty (CMC)
Range
1,5 and 10MHz
Uncertainty (CMC)
1,0E-9 * f

Remarks/ref. to internal procedure: BL21.04.03

Frequency, as measured quantity

Range
Uncertainty (CMC)
Range
0,5Hz to 5GHz
Uncertainty (CMC)
1,0E-9*f

Remarks/ref. to internal procedure: BL21.04.03

Frequency, calibration of meter

Range
Uncertainty (CMC)
Range
0,010Hz to 2,0MHz
Uncertainty (CMC)
5,0 µHz + 1,7 ppm

Remarks/ref. to internal procedure: BL08.04.01

Frequency, calibration of source

Range
Uncertainty (CMC)
Range
3Hz to 1 MHz
Uncertainty (CMC)
50 µHz + 50 ppm

Remarks/ref. to internal procedure: BL08.05.01

Time interval

Range
Uncertainty (CMC)
Range
1 to 200 hour
Uncertainty (CMC)
0,80s

Remarks/ref. to internal procedure: BL21.04.04

Number, number of pulses

Range
Uncertainty (CMC)
Range
1 to 10G
Uncertainty (CMC)
0,60

Remarks/ref. to internal procedure: BL21.04.06
The administrative / geographical unit:
IKM Laboratorium avd Oslo, field calibration
Ensjøveien 18
0661 Oslo



Field testing laboratory K01 Electrical quantities

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,20 %

Remarks/ref. to internal procedure: BL08.07.02 cosφ = -1 til 1 f = 45-70Hz I = 0,01-100A U = 5-520V

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,040 %

Remarks/ref. to internal procedure: BL08.07.03 Based on "Statnetts krav til måling av sentralnettsutveksling v.1.1(2016)" and includes evaluation of metering point. cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V

Energy (and power) Active, Reactive, Apparent 1 and 3 phase

Range
Uncertainty (CMC)
Range
See remark
Uncertainty (CMC)
0,020 %

Remarks/ref. to internal procedure: BL08.07.04 cosφ = -1 til 1 f = 45-70Hz I = 0,01-120A U = 5-520V