You are here:
Norwegian accreditation

Accreditation scope for

CAL 007

Kiwa AS

Kalibrering

Postboks 141 Økern
0509 Oslo

Telephone: +47 22 86 50 00
E-mail: firmapost@kiwa.com
Internet: http://www.kiwa.no
The laboratory meets the requirements in

NS-EN ISO/IEC 17025:2017




Accreditation includes:
The administrative / geographical unit:
Kiwa AS
Bleivassvegen 28B,
5347 Ågotnes



Permanent facility K01 Electrical quantities

Resistance (Dekade resistors and fixed resistors)

Range
Uncertainty (CMC)
Range
0 - 12 Ω
Uncertainty (CMC)
10ppm x R + 30µΩ
Range
12 - 120 Ω
Uncertainty (CMC)
7,0ppm x R + 80 µΩ
Range
120 - 1200 Ω
Uncertainty (CMC)
8,0ppm x R + 80 µΩ
Range
1,2 - 12 kΩ
Uncertainty (CMC)
6,2ppm x R
Range
12 - 120 kΩ
Uncertainty (CMC)
10ppm x R
Range
120 - 1200 kΩ
Uncertainty (CMC)
13ppm x R
Range
1,2 - 12 MΩ
Uncertainty (CMC)
36ppm x R
Range
12 - 120 MΩ
Uncertainty (CMC)
500ppm x R

Remarks/ref. to internal procedure: KH 2-9-1 AC, DC resistance (by measurement), (​​R = measured value)

Resistance

Range
Uncertainty (CMC)
Range
0,2 - 0,9 Ω
Uncertainty (CMC)
3,5ppm x R
Range
0,9 - 1,1 Ω
Uncertainty (CMC)
2,1ppm x R
Range
1,1 - 3 Ω
Uncertainty (CMC)
2,5ppm x R
Range
3 - 9 Ω
Uncertainty (CMC)
3,7ppm x R
Range
9 - 11 Ω
Uncertainty (CMC)
3,1ppm x R
Range
11 - 30 Ω
Uncertainty (CMC)
3,4ppm x R
Range
30 - 90 Ω
Uncertainty (CMC)
2,1ppm x R
Range
90 - 110 Ω
Uncertainty (CMC)
1,6ppm x R
Range
110 - 300 Ω
Uncertainty (CMC)
2,1ppm x R
Range
300 - 900 Ω
Uncertainty (CMC)
3,4ppm x R
Range
900 - 1100 Ω
Uncertainty (CMC)
3,0ppm x R
Range
1,1 - 3 kΩ
Uncertainty (CMC)
3,4ppm x R
Range
3 - 9 kΩ
Uncertainty (CMC)
2,5ppm x R
Range
9 - 11 kΩ
Uncertainty (CMC)
2,1ppm x R
Range
11 - 30 kΩ
Uncertainty (CMC)
3,5ppm x R
Range
30 - 90 kΩ
Uncertainty (CMC)
4,0ppm x R
Range
90 - 110 kΩ
Uncertainty (CMC)
3,5ppm x R

Remarks/ref. to internal procedure: KH-2-9-1 BB, DC resistance (by measurement), (​​R = measured value)

DC- Voltage

Range
Uncertainty (CMC)
Range
0 V
Uncertainty (CMC)
50 nV (short)
Kortslutning
Range
0 mV - 130 mV
Uncertainty (CMC)
0,9 µV
Range
130 mV - 500 mV
Uncertainty (CMC)
6,9ppm x U
Range
500 mV - 2,2 mV
Uncertainty (CMC)
4,4ppm x U
Range
2,2 V - 11 V
Uncertainty (CMC)
3,4ppm x U
Range
11 V - 22 V
Uncertainty (CMC)
3,7ppm x U
Range
22 V - 220 V
Uncertainty (CMC)
4,0ppm x U
Range
220 V - 1100 V
Uncertainty (CMC)
4,0ppm x U

Remarks/ref. to internal procedure: KH-S 2-9-1-S, DC voltage (as the source), (​​U = measured value)

DC Power

Range
Uncertainty (CMC)
Range
0 A (Open)
Uncertainty (CMC)
0,5 nA
Range
0 µA - 220 µA
Uncertainty (CMC)
14ppm x I + 4 nA
Range
220 µA - 2,2 mA
Uncertainty (CMC)
13ppm x I
Range
2,2 mA - 22 mA
Uncertainty (CMC)
12ppm x I
Range
22 mA - 100 mA
Uncertainty (CMC)
12ppm x I
Range
100 mA - 220 mA
Uncertainty (CMC)
14ppm x I
Range
220 mA - 1 mA
Uncertainty (CMC)
16ppm x I
Range
1 A - 2,2 A
Uncertainty (CMC)
32ppm x I
Range
2,2 A - 11A
Uncertainty (CMC)
330ppm x I

Remarks/ref. to internal procedure: KH 2-9-1-T, DC power (as the source), (​​I = measured value)

DC Voltage

Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
0 - 0,12 V
Uncertainty (CMC)
6,4ppm x U + 0,32 µV
Range
0,12 - 1,2 V
Uncertainty (CMC)
3,4ppm x U + 0,6 µV
Range
1,2 - 12 V
Uncertainty (CMC)
2,9ppm x U + 0,9 µV
Range
12 - 120 V
Uncertainty (CMC)
5,3ppm x U + 30uV
Range
120 - 450 V
Uncertainty (CMC)
9,0ppm x U
Range
450 - 700 V
Uncertainty (CMC)
13ppm x U
Range
700 - 1000 V
Uncertainty (CMC)
17ppm x U

Remarks/ref. to internal procedure: KH-2-9-1-AP, DC voltage (by measurement), (​​U = measured value)

AC voltage (as the source)

Range
Uncertainty (CMC)
Range
0,22mV - 2,2mV:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
0,30% x U + 4,5 µV
Range
20 - 40Hz
Uncertainty (CMC)
0,30% x U + 4,5 µV
Range
40 - 20000Hz
Uncertainty (CMC)
0,29% x U + 4,5 µV
Range
20 - 50kHz
Uncertainty (CMC)
0,30% x U + 4,5 µV
Range
50 - 100kHz
Uncertainty (CMC)
0,32% x U + 7,0 µV
Range
100 - 300kHz
Uncertainty (CMC)
0,60% x U + 13 µV
Range
300 - 500kHz
Uncertainty (CMC)
1,1% x U + 25 µV
Range
500 - 1000kHz
Uncertainty (CMC)
1,4% x U + 25 µV
Range
2,2mV - 22mV:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
660ppm x U + 5,0 µV
Range
20 - 40Hz
Uncertainty (CMC)
470ppm x U + 5,0 µV
Range
40 - 20000Hz
Uncertainty (CMC)
510ppm x U + 5,0 µV
Range
20 - 50kHz
Uncertainty (CMC)
600ppm x U + 5,0 µV
Range
50 - 100kHz
Uncertainty (CMC)
0,11% x U + 7,0 µV
Range
100 - 300kHz
Uncertainty (CMC)
0,16% x U + 12 µV
Range
300 - 500kHz
Uncertainty (CMC)
0,30% x U + 25 µV
Range
500 - 1000kHz
Uncertainty (CMC)
0,73% x U + 25 µV
Range
22mV - 200mV:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
520ppm x U + 13 µV
Range
20 - 40Hz
Uncertainty (CMC)
240ppm x U + 8,0 µV
Range
40 - 20000Hz
Uncertainty (CMC)
160ppm x U + 8,0 µV
Range
20 - 50kHz
Uncertainty (CMC)
310ppm x U + 8,0 µV
Range
50 - 100kHz
Uncertainty (CMC)
770ppm x U + 8,0 µV
Range
100 - 300kHz
Uncertainty (CMC)
0,12% x U + 25 µV
Range
300 - 500kHz
Uncertainty (CMC)
0,22% x U + 35 µV
Range
500 - 1000kHz
Uncertainty (CMC)
0,66% x U + 80 µV
Range
0,22V - 2,2V:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
470ppm x U + 80 µV
Range
20 - 40Hz
Uncertainty (CMC)
190ppm x U + 25 µV
Range
40 - 20000Hz
Uncertainty (CMC)
90ppm x U + 6,0 µV
Range
20 - 50kHz
Uncertainty (CMC)
150ppm x U + 16 µV
Range
50 - 100kHz
Uncertainty (CMC)
260ppm x U + 70 µV
Range
100 - 300kHz
Uncertainty (CMC)
560ppm x U + 130 µV
Range
300 - 500kHz
Uncertainty (CMC)
0,15% x U + 350 µV
Range
500 - 1000kHz
Uncertainty (CMC)
0,38% x U + 850 µV
Range
2,2V - 22V:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
470ppm x U + 800 µV
Range
20 - 40Hz
Uncertainty (CMC)
190ppm x U + 250 µV
Range
40 - 20000Hz
Uncertainty (CMC)
110ppm x U + 60 µV
Range
20 - 50kHz
Uncertainty (CMC)
140ppm x U + 160 µV
Range
50 - 100kHz
Uncertainty (CMC)
260ppm x U + 350 µV
Range
100 - 300kHz
Uncertainty (CMC)
620ppm x U + 1,5 mV
Range
300 - 500kHz
Uncertainty (CMC)
0,16% x U + 4,3 mV
Range
500 - 1000kHz
Uncertainty (CMC)
0,40% x U + 8,5 mV
Range
22V - 220V:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
470ppm x U + 8,0 mV
Range
20 - 40Hz
Uncertainty (CMC)
190ppm x U + 2,5 mV
* Spenningen U er
Range
40 - 20000Hz
Uncertainty (CMC)
110ppm x U + 800 µV
begrenset slik at
Range
20 - 50kHz
Uncertainty (CMC)
210ppm x U + 3,5 mV
fxU ikke er mer enn:
Range
50 - 100kHz
Uncertainty (CMC)
480ppm x U + 8,0 mV
2,2 x 10-7 V x Hz
Range
220V - 1100V:
Uncertainty (CMC)
Range
40 - 1000Hz
Uncertainty (CMC)
130ppm x U + 4,0 mV
Range
1 - 20kHz
Uncertainty (CMC)
200ppm x U + 6,0 mV

Remarks/ref. to internal procedure: Procedure: KH-U 2-9-1, AC voltage (the source), (​​U = nominal value)

AC power (as the source)

Range
Uncertainty (CMC)
Range
9uA - 220uA:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
650ppm x I + 25 nA
Range
20 - 40Hz
Uncertainty (CMC)
350ppm x I + 20 nA
Range
40 - 1000Hz
Uncertainty (CMC)
190ppm x I + 16 nA
Range
1 - 5kHz
Uncertainty (CMC)
0,10% x I + 40 nA
Range
5 - 10kHz
Uncertainty (CMC)
0,27% x I + 80 nA
Range
0,22mA - 2,2mA:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
650ppm x I + 40 nA
Range
20 - 40Hz
Uncertainty (CMC)
340ppm x I + 35 nA
Range
40 - 1000Hz
Uncertainty (CMC)
180ppm x I + 35 nA
Range
1 - 5kHz
Uncertainty (CMC)
880ppm x I + 400 nA
Range
5 - 10kHz
Uncertainty (CMC)
0,20% x I + 800 nA
Range
2,2mA - 22mA:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
610ppm x I + 400 nA
Range
20 - 40Hz
Uncertainty (CMC)
330ppm x I + 350 nA
Range
40 - 1000Hz
Uncertainty (CMC)
160ppm x I + 350 nA
Range
1 - 5kHz
Uncertainty (CMC)
0,13% x I + 4,0 uA
Range
5 - 10kHzz
Uncertainty (CMC)
0,27% x I + 8,0 uA
Range
22mA - 220mA:
Uncertainty (CMC)
Range
10 - 20Hz
Uncertainty (CMC)
610ppm x I + 4,0 uA
Range
20 - 40Hz
Uncertainty (CMC)
330ppm x I + 3,5 uA
Range
40 - 1000Hz
Uncertainty (CMC)
160ppm x I + 3,5 uA
Range
1 - 5kHz
Uncertainty (CMC)
0,13% x I + 40 uA
Range
5 - 10kHz
Uncertainty (CMC)
0,27% x I + 80 uA
Range
0,22A - 2,2A:
Uncertainty (CMC)
Range
40 - 1000Hz
Uncertainty (CMC)
570ppm x I + 35 uA
Range
1 - 5kHz
Uncertainty (CMC)
0,41% x I + 80 uA
Range
5 - 10kHz
Uncertainty (CMC)
1,5% x I + 160 uA
Range
2,2A - 11A:
Uncertainty (CMC)
Range
40 - 1000Hz
Uncertainty (CMC)
500ppm x I + 170 uA
Range
1 - 5kHz
Uncertainty (CMC)
0,23% x I + 380 uA
Range
5 - 10kHz
Uncertainty (CMC)
0,78% x I + 580 uA

Remarks/ref. to internal procedure: KH 2-9-1-W AC power (as the source), (​​I = nominal value)

DC Power

Range
Uncertainty (CMC)
Range
0 mA - 0,1 mA
Uncertainty (CMC)
3,8nA
Range
0,1 mA - 0,5 mA
Uncertainty (CMC)
6,2nA
Range
0,5 mA - 1,2 mA
Uncertainty (CMC)
10nA
Range
1,2 mA - 2,2 mA
Uncertainty (CMC)
15nA
Range
2,2 mA - 5,0 mA
Uncertainty (CMC)
25nA
Range
5,0 mA - 7,0 mA
Uncertainty (CMC)
33nA
Range
7,0 mA - 12 mA
Uncertainty (CMC)
44nA
Range
12 mA - 14 mA
Uncertainty (CMC)
51nA
Range
14 mA - 22 mA
Uncertainty (CMC)
75nA
Range
22 mA - 40 mA
Uncertainty (CMC)
0,13µA
Range
40 mA - 70 mA
Uncertainty (CMC)
0,22µA
Range
70 mA - 100 mA
Uncertainty (CMC)
0,31µA
Range
100 mA - 120 mA
Uncertainty (CMC)
1,0µA
Range
0,12 A - 0,22 A
Uncertainty (CMC)
1,5µA
Range
0,22 A - 0,50 A
Uncertainty (CMC)
3,0µA
Range
0,50 A - 1,0 A
Uncertainty (CMC)
130µA
Range
1 A - 6 A
Uncertainty (CMC)
0,60mA
Range
6 A - 9,9 A
Uncertainty (CMC)
1,0mA
Range
9,9 A - 10,1 A
Uncertainty (CMC)
0,8mA
Range
10,1 A - 19,9 A
Uncertainty (CMC)
8,0mA
Range
19,9 A - 20 A
Uncertainty (CMC)
1,5mA
Range
20 A - 30 A
Uncertainty (CMC)
8,0mA
Range
30 A - 50 A
Uncertainty (CMC)
38mA

Remarks/ref. to internal procedure: KH-2-9-1-AN, DC power (by measurement), (​​I = measured value)

DC resistance (as sources)

Range
Uncertainty (CMC)
Range
0 Ω
Uncertainty (CMC)
5,0 µΩ
Range
0,01 Ω
Uncertainty (CMC)
0,9 µΩ
Range
0,02 Ω
Uncertainty (CMC)
1,9 µΩ
Range
1 Ω
Uncertainty (CMC)
3,0 µΩ
Range
1,9 Ω
Uncertainty (CMC)
38 µΩ
Range
10 Ω
Uncertainty (CMC)
30 µΩ
Range
19 Ω
Uncertainty (CMC)
140 µΩ
Range
100 Ω
Uncertainty (CMC)
210 µΩ
Range
190 Ω
Uncertainty (CMC)
1,3 mΩ
Range
1000 Ω
Uncertainty (CMC)
3,0 mΩ
Range
1900 Ω
Uncertainty (CMC)
11 mΩ
Range
10 kΩ
Uncertainty (CMC)
21 mΩ
Range
19 kΩ
Uncertainty (CMC)
130 mΩ
Range
100 kΩ
Uncertainty (CMC)
220 mΩ
Range
190 kΩ
Uncertainty (CMC)
1,6 Ω
Range
1 MΩ
Uncertainty (CMC)
7 Ω
Range
1,9 MΩ
Uncertainty (CMC)
28 Ω
Range
10 MΩ
Uncertainty (CMC)
170 Ω
Range
19 MΩ
Uncertainty (CMC)
420 Ω
Range
100 MΩ
Uncertainty (CMC)
3,6 kΩ

Remarks/ref. to internal procedure: KH-2-9-1
Permanent facility K03 Time and frequency

Frequency (measurement)

Range
Uncertainty (CMC)
Range
1 MHz
Uncertainty (CMC)
7,5 x 10-12 x f
Range
5 MHz
Uncertainty (CMC)
7,5 x 10-12 x f
Range
10 MHz
Uncertainty (CMC)
7,5 x 10-12 x f

Remarks/ref. to internal procedure: KH 2-9-1 AK

Frequency (measurement with frequency counter)

Range
Uncertainty (CMC)
Range
1 Hz til/to 10 Hz
Uncertainty (CMC)
1,0 x 10-8 x f
Square Sign
Range
10 Hz til/to 1,3 GHz
Uncertainty (CMC)
1,0 x 10-9 x f
Square Sign
Range
1 Hz til/to 2 Hz
Uncertainty (CMC)
1,0 x 10-6 x f
Sine Signal
Range
2 Hz til/to 20 Hz
Uncertainty (CMC)
1,0 x 10-8 x f
Sine Signal
Range
20 Hz til/to 1,3 GHz
Uncertainty (CMC)
1,0 x 10-9 x f
Sine Signal

Remarks/ref. to internal procedure: KH 2-9-1 AK

Frequency (generation)

Range
Uncertainty (CMC)
Range
1 µHz til/to 4 GHz
Uncertainty (CMC)
1,0 x 10-10 x f

Remarks/ref. to internal procedure: KH 2-9-1 AH
Permanent facility K05 Mechanical quantities

Mass (weight)

Range
Uncertainty (CMC)
Range
Fra 0,001 g til og med 20 g
Uncertainty (CMC)
0,10 mg
Range
Fra 20 g til og med 205 g
Uncertainty (CMC)
0,16 mg
Range
Fra 205 g til og med 500 g
Uncertainty (CMC)
0,4 mg
Range
Fra 500 g til og med 1100 g
Uncertainty (CMC)
0,5 mg
Range
Fra 1,1 kg til og med 10 kg
Uncertainty (CMC)
25 mg
Range
Fra 10 kg til og med 20 kg
Uncertainty (CMC)
60 mg

Remarks/ref. to internal procedure: KH 2-9-1 AV, ID-1800

Pressure (gauge)

Range
Uncertainty (CMC)
Range
-100 kPa til/to 0 hPa
Uncertainty (CMC)
7,8 til/to 0,25 Pa
Range
0 hPa til/to 100 kPa
Uncertainty (CMC)
0,25 til/to 7 Pa
Range
100 kPa til/to 2,5MPa
Uncertainty (CMC)
7 til/to 130 Pa
Range
2,5 MPa til/to 25 Mpa
Uncertainty (CMC)
0,13 til/to 1,2 kPa
Range
25 MPa til/to 100 Mpa
Uncertainty (CMC)
1,2 til/to 5,6 kPa
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: KH 2-9-1 A, KH 2-9-1 AG

Pressure (abs)

Range
Uncertainty (CMC)
Range
0 hPa til/to 1200 hPa
Uncertainty (CMC)
3,5 Pa til/to 10 Pa
Range
1200 hPa til/to 200 kPa
Uncertainty (CMC)
10 Pa til/to 17 Pa
Range
200 kPa til/to 2,6 Mpa
Uncertainty (CMC)
17 Pa til/to 140 Pa
Range
2,6 MPa til/to 25 Mpa
Uncertainty (CMC)
0,14 kPa til/to 1,2 kPa
Range
25 MPa til/to 100 Mpa
Uncertainty (CMC)
1,2 kPa til/to 5,6 kPa
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: KH 2-9-1 A, KH 2-9-1 AG

Piston Area oil

Range
Uncertainty (CMC)
Range
40 MPa til/to 100 Mpa
Uncertainty (CMC)
55 ppm
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: KH 2-9-1 G

Pressure (diff)

Range
Uncertainty (CMC)
Range
0 Mpa til/to 0,4 Mpa
Uncertainty (CMC)
2 Pa til/to 24 Pa
Range
0,4 Mpa til/to 20 Mpa
Uncertainty (CMC)
24 Pa til/to 970 Pa

Remarks/ref. to internal procedure: KH 2-9-1 A, KH 2-9-1 AG

Piston Area gas

Range
Uncertainty (CMC)
Range
15 hPa til/to 100 kPa
Uncertainty (CMC)
75 ppm
Range
100 kPa til/to 2,5MPa
Uncertainty (CMC)
55 ppm
Range
2,5 MPa til/to 25 Mpa
Uncertainty (CMC)
48 ppm
Range
Uncertainty (CMC)
Range
Uncertainty (CMC)

Remarks/ref. to internal procedure: KH 2-9-1 AE
Permanent facility K07 Fluid quantities

Density of gas (kg/m3 in the pressure range)

Range
Uncertainty (CMC)
Range
1 kg/m3 til/to130 kg/m3
Uncertainty (CMC)
0,5 g/m3 + 0,028 % av avlest
Range
130 kg/m3 til/to 300 kg/m3
Uncertainty (CMC)
0,052 % av avlest

Remarks/ref. to internal procedure: KH 2-9-1 Q (ID-3695) og KH 2-9-1 M (ID-1804), 270 to 350 ° K, Nitrogen

Density of liquids

Range
Uncertainty (CMC)
Range
650 kg/m3 til/to 1350 kg/m3
Uncertainty (CMC)
80 g/m3 til/to 100 g/m3

Remarks/ref. to internal procedure: KH 2-9-1 N and V, 10 °C to 40 °C
Permanent facility K10 Temperature, humidity and thermophysical properties

Direct reading instruments

Range
Uncertainty (CMC)
Range
+200 °C til/to +280 °C
Uncertainty (CMC)
17mK
Range
-80 °C til/to -50 °C
Uncertainty (CMC)
20mK
Range
-50 °C til/to -10 °C
Uncertainty (CMC)
15 mK
Range
-10 °C til/to +100 °C
Uncertainty (CMC)
10mK
Range
+100 °C til/to +200 °C
Uncertainty (CMC)
15mK

Remarks/ref. to internal procedure: KH 2-9-1 AW

Resistance thermometers

Range
Uncertainty (CMC)
Range
+200 °C til/to +280 °C
Uncertainty (CMC)
17 mK
Range
-80 °C til/to -50 °C
Uncertainty (CMC)
20 mK
Range
-50 °C til/to -10 °C
Uncertainty (CMC)
15 mK
Range
-10 °C til/to+100 °C
Uncertainty (CMC)
10 mK
Range
+100 °C til/to +200 °C
Uncertainty (CMC)
15 mK

Remarks/ref. to internal procedure: KH 2-9-1 F

Temperature transmitters

Range
Uncertainty (CMC)
Range
-80 °C til/to -50 °C
Uncertainty (CMC)
30mK
Range
-50 °C til/to -10 °C
Uncertainty (CMC)
25mK
Range
-10 °C til/to 100 °C
Uncertainty (CMC)
20mK
Range
100 °C til/to 280 °C
Uncertainty (CMC)
25mK

Remarks/ref. to internal procedure: KH 2-9-1 AY
The administrative / geographical unit:
Kiwa AS
Kabelgaten 2
0580 Oslo



Field testing laboratory K05 Mechanical quantities

Brake pressure sensor. Manometer

Range
Uncertainty (CMC)
Range
0 bar til/to 16 bar
Uncertainty (CMC)
± 0,08 bar

Remarks/ref. to internal procedure: ID-1989 Analog reading

Brake pressure sensor. Manometer

Range
Uncertainty (CMC)
Range
0 bar til/to 16 bar
Uncertainty (CMC)
± 0,06 bar

Remarks/ref. to internal procedure: ID-1989 Digital reading

Deceleration measurement

Range
Uncertainty (CMC)
Range
0-100 %g
Uncertainty (CMC)
± 1,3%g v/28%g = 0,13 m/s2 ± 1,3%g v/60%g = 0,13 m/s2
Bowmonk (digital)
Range
0-100 %g
Uncertainty (CMC)
±0,56%g v/28%g = 0,06 m/s2 ±0,57%g v/60%g = 0,06 m/s2
MotoMeter
Range
0-6,00 m/s2
Uncertainty (CMC)
±0,02 m/s2 v/2,8 m/s2 ±0,03 m/s2 v/6,0 m/s2
MaHa (digital)
Range
0-100 %g
Uncertainty (CMC)
± 1,2%g v/30%g = 0,12 m/s2 ± 1.2%g v/75%g = 0,12 m/s2
Bowmonk (analog)

Remarks/ref. to internal procedure: ID-1990

Brake tester - Dead weight

Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
0,063 kN v/8 kN
Range
1kN til/to 8 kN (personbil)
Uncertainty (CMC)
± 0,024 kN v/ 1 kN

Remarks/ref. to internal procedure: ID-1986

Brake tester - Load cell

Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
0,061 v/6 kN
Range
1 kN til 6 kN (personbil)
Uncertainty (CMC)
± 0,024 kN v/ 1 kN

Remarks/ref. to internal procedure: ID-1986
Field testing laboratory K07 Fluid quantities

Exhaust gases analyser - Hydrocarbons

Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
25 - 1750 vol ppm Hexan
Uncertainty (CMC)
± 2 vol ppm v/ 25 vol ppm Hexan ± 18 vol ppm v/ 1750 vol vol ppm Hexan

Remarks/ref. to internal procedure: ID-1985 Calibrations with two gas concentrations

Exhaust gases analyser - Carbon dioxide (CO2)

Range
Uncertainty (CMC)
Range
6,5 - 14,00 vol%
Uncertainty (CMC)
± 0,07 vol% v/ 6,5 vol% ± 0,15 vol% v/ 14,0 vol%

Remarks/ref. to internal procedure: ID-1985 Calibration with two gas concentrations

Exhaust gases analyser - Carbon dioxide (CO)

Range
Uncertainty (CMC)
Range
0,5 - 4,0 vol%
Uncertainty (CMC)
± 0,02 vol% v/ 0,5 vol% ± 0,04 vol% v/ 4,0 vol%

Remarks/ref. to internal procedure: ID-1985 Calibration with two gas concentrations

Exhaust gases - Oxygen (O2)

Range
Uncertainty (CMC)
Range
Uncertainty (CMC)
Range
0,2 vol% - 21 vol%
Uncertainty (CMC)
± 0,04 vol% v/2,0 vol% ± 0,07 vol% v/0,2 vol%

Remarks/ref. to internal procedure: ID-1985 Calibrations with two gas concentrations

Diesel smoke meter - k-value

Range
Uncertainty (CMC)
Range
0,8 - 4,0 m-1
Uncertainty (CMC)
± 0,09 m-1 v/ 0,8 m-1 ± 0,10 m-1 v/ 3,7 m-1

Remarks/ref. to internal procedure: ID-1991 Calibration with two or more filters

Exhaust gases analyser - Lambda (Calculated range based on the stated gas mixtures).

Range
Uncertainty (CMC)
Range
0,8 - 1,2
Uncertainty (CMC)
± 0,002 v/1,155 ± 0,003 v/0,870

Remarks/ref. to internal procedure: ID-1985 Calibration with two gas concentrations
Field testing laboratory K08 Optical quantities

Headlamp beam tester

Range
Uncertainty (CMC)
Range
0 - 40 mm/m nedvinkling
Uncertainty (CMC)
± 1,2 mm/m

Remarks/ref. to internal procedure: ID-1993