Direct reading insturments
Range
-80 °C til -50 °C
Uncertainty (CMC)
± 20 mK
Range
-50 °C til 100 °C
Uncertainty (CMC)
± 15 mK
Range
100 °C til 200 °C
Uncertainty (CMC)
± 20 mK
Range
200 °C til 400 °C
Uncertainty (CMC)
± 40 mK
Range
400 °C til 500 °C
Uncertainty (CMC)
± 65 mK
Range
500 °C til 660 °C
Uncertainty (CMC)
± 0,11 K
Remarks/ref. to internal procedure: Reference instrument: SPRT/Quartz thermometer.
Method 010.
Resistance thermometers
Range
- 80 °C til - 50 °C
Uncertainty (CMC)
± 20 mK
Range
- 50 °C til 100 °C
Uncertainty (CMC)
± 15 mK
Range
100 °C til 200 °C
Uncertainty (CMC)
± 20 mK
Range
200 °C til 400 °C
Uncertainty (CMC)
± 40 mK
Range
400 °C til 500 °C
Uncertainty (CMC)
± 65 mK
Range
500 °C til 660 °C
Uncertainty (CMC)
± 0,11 K
Remarks/ref. to internal procedure: Reference instrument: SPRT/Quartz thermometer.
Method 012.
Thermcouples (types K, T, J, N, E, R, S and B)
Range
- 80 °C til - 40 °C
Uncertainty (CMC)
± 0,4 K
Range
- 40 °C til 300 °C
Uncertainty (CMC)
± 0,3 K
Range
300 °C til 500 °C
Uncertainty (CMC)
± 0,4 K
Range
500 °C til 660 °C
Uncertainty (CMC)
± 0,6 K
Remarks/ref. to internal procedure: Reference instrument: SPRT/Quartz thermometer.
Method 006.
Temperature transmitters
Range
0 °C til 100 °C
Uncertainty (CMC)
± 22 mK
Remarks/ref. to internal procedure: Reference instrument: SPRT/Quartz thermometer.
Metode 013.
Valid for instruments with measurement range from -0 °C to 100 °C.
Temperature transmitters
Range
- 50 °C til 100 °C
Uncertainty (CMC)
± 40 mK
Range
100 °C til 250 °C
Uncertainty (CMC)
± 60 mK
Remarks/ref. to internal procedure: Referense instrument: SPRT/Quartz thermometer.
Metode 013.
Valid for instruments with measurement range from -50 °C to 250 °C.
Temperature transmitters
Range
- 80 °C til 100 °C
Uncertainty (CMC)
± 75 mK
Range
100 °C til 300 °C
Uncertainty (CMC)
± 95 mK
Range
300 °C til 500 °C
Uncertainty (CMC)
± 0,12 K
Range
500 °C til 660 °C
Uncertainty (CMC)
± 0,16 K
Remarks/ref. to internal procedure: Reference instrument: SPRT/Quartz thermometer.
Metode 013.
Valid for instruments with measurement range from - 80 °C to 660 °C.
Logs with built-in sensor in climatic chamber
Range
- 40 °C til - 20 °C
Uncertainty (CMC)
± 0,81 K
Range
- 20 °C til 0 °C
Uncertainty (CMC)
± 0,7 K
Range
0 °C til 25 °C
Uncertainty (CMC)
± 0,18 K
Range
25 °C til 50 °C
Uncertainty (CMC)
± 0,22 K
Range
50 °C til 100 °C
Uncertainty (CMC)
± 0,45 K
Range
100 °C til 150 °C
Uncertainty (CMC)
± 0,77 K
Remarks/ref. to internal procedure: Testo 735 with PT100.
Method 008.